- Chamber volume: NC 150: 162 litres / NC 150D: 150 litres.
- Sterilization temperatures: 121°C, 125°C, 134°C, 136°C.
- Advanced technology for the sterilization of textile, wrapped or packed materials, glass and liquid.
- With its double doors, NC 150D is the ideal solution for the small sterilization centres which are divided as clean and sterile areas.
- Memory for the last 100 failures.
- 7 preset programs: Resistant textile, Wrapped porous, Prion, Quick Unwrapped, Resistant rubber, Sensitive material, Gravity.
- 10 free programs and two special programs for liquid sterilization and drying.
- Vacuum leakage and Bowie & Dick test programs.
- Panel type printer to keep the records of the sterilization cycle.
- USB port to record the operated cycles in the memory to a memory stick.
- HEPA filter at air intake line.
- Massive data storage with internal memory which stores the records of last 500 cycles in detail and 29.500 cycles as summary.
- Colour of the display changes to red in case of any alarm condition along with an audible alarm.
- Pre-heated chamber to eliminate condensation on the chamber walls and speed up the heating of the samples.
- Integrated steam generator for fast production of steam.
- Maintenance-free diaphragm type powerful vacuum pump.
- Air driven gasket by with internal compressor.
- Option to use with shelf or basket for easy loading
- Fractionated pre-vacuum to eliminate all air bubbles and achieve successful steam penetration.
- Fractionated post-vacuum for efficient drying of the samples.
- Pressure measurement of gasket.
- Ethernet port for remote access through internet by means of optional NuveCloser™ software.
- Comprehensive self – diagnostic system for possible failures.
- Powerful and reliable N-Smart™ microprocessor control system.
- Wi-Fi connection for N-Mobile™ application for traceability and instant notifications
- Operating parameters can be tracked graphically against time.
- Process evaluation system.
- Sending e-mails up to five e-mail addresses with the details of failure.
- Password protected control system, service and calibration menu.
- Reminder for gasket and filter replacement.
- Remote failure diagnostics by means of NuveCloser™ software.
- Sterilization quality matching the needs of EN 285 European Standard for Large Steam Sterilizers.
- Optional AlerText™ SMS alarm system for ultimate security.
- Safety valve against over pressure.
- Certified chamber according to PED 2014/68 EU.